Comparator:state - Teledyne T3LCR1002 User Manual

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

The COMParator subsystem command group sets the comparator function,
including its ON/OFF setting, limit mode, and limit values.
The COMParator:STATe command sets the comparator function to ON or
OFF.
1.800.561.8187
COMParator
Command
COMParator:STATe {ON,OFF,1,0}
Syntax
Parameter
Where,
ON or 1 Sets the comparator to ON
OFF or 0 Sets the comparator to OFF
Example
SEND> COMP:STAT OFF<NL>
Query Syntax

COMParator:STATe?

Query Response
{on,off}
Example
SEND> COMP:STAT?<NL>
RET>
www.
:S TATe
{ON,OFF}
:MODE
{ABS,PER,SEQ}
:AUX
{ON,OFF}
:BINS
{1 to 9}
:TOLerance
:NOMinal
:BIN
:S ECONDARY(S LIM)
:BEEP
{OFF,PASS,FAIL}
:OPEN
{OFF,2,5,10,20,50}
on<NL>
.com
T3LCR Series User Manual
<value>
<n,low,high>
<low>,<high>
113