Only four measurement values will be shown in this
simplified display mode: The primary measurement
parameters, the secondary measurement parameters and
two monitored parameters such as , %, Iac or Vac,
etc.
Press the ENLARGE soft-key in the [MEAS DISPLAY]
to enter the [ENLARGE DISPLAY] mode (simplified
display mode); on the contrary, press the NORMAL
soft-key in the [ENLARGE DISPLAY] mode to quit the
[ENLARGE DISPLAY] mode and return to the regular
[MEAS DISPLAY] mode.
Two biggest readings you see in the upper part of the
[ENLARGE DISPLAY] mode are the primary and the
secondary measurement parameters while the two other
smaller readings in the lower part of the display are the
monitored parameters you chose to be displayed. The
PASS/FAIL result of the Direct Comparison function
(see 4.5.2 Direct Comparison Function) for the DUT
currently being measured will be displayed in the bottom
right corner of the enlarged display.
The Direct Comparison function allows you to see
whether the currently measured DUT is within the
acceptable tolerance range without the complexity of the
BIN sorting function.