Table of Contents

Advertisement

3-2 Turning ON the main switch

16
When using the reflected LED light source (BX3M-LEDR)
Set the main switch a to (ON).
1
When using the light source other than the reflected LED
light source (BX3M-LEDR)
1
Set the main switch of each unit to (ON).
a
* When using the mercury burner, the arc image becomes stable 5
TIP
NOTE
When using the control box (BX53M-CBFM)
1
Set the main switch to (ON).
TIP
Combinable units
Main switch of the power supply for
Mercury burner
mercury burner (U-RFL-T)*
Main switch of the power supply for
Halogen bulb
halogen bulb (TH4-100/TH4-200)
Power supply switch of the LED and
Light source
LDP light source (U-LGPS)
to 10 minutes after turning it ON.
The discharge type mercury burner may not be turned ON
in one operation due to its nature. In this case, set the main
switch to
 (OFF) once, wait 5 to 10 seconds and set it to
 (ON) again.
· Do not turn OFF the mercury burner for two hours after
turning it ON in order to prevent the burner from reducing
its lifetime.
· When turning ON the mercury burner again, wait until
the vapor of the mercury burner is cooled down and
devolatilized. Wait approx. 10 minutes and turn it ON.
The buzzer sound is heard once.
Unit to set to ON