HP VXI 75000 C Series User's And Scpi Programming Manual page 338

Algorithmic closed loop controller
Table of Contents

Advertisement

338 Error Messages
3052
FIIFO Value
1 - 99
100 - 163
164
200
201
202
203
204
Test ID
1 - 19, 21 - 29
20, 30 -37
38 - 71
72,74 - 76, 80 - 93,
301 - 354
73, 77 - 79, 94 - 99
NOTE
During the first 5 minutes after power is applied, *TST? may fail. Allow
the module to warm-up before executing *TST?
'Self test failed. Test info in FIFO'. Use
SENS:DATA:FIFO:ALL? to retrieve data from FIFO.
NOTE: *TST? always sets the FIFO data FORMat to
ASCII,7. Read FIFO data into string variables.
Definition
ID number of failed test (see following table for
possible corrective actions)
channel number(s) associated with test (ch 0-63)
special "channel" used for A/D tests only
A/D range 0.0625V associated with failed test
A/D range 0.25V associated with failed test
A/D range 1V associated with failed test
A/D range 4V associated with failed test
A/D range 16V associated with failed test
Corrective Actions
(HP Service)*
Remove all SCPs and see if *TST? passes. If so,
replace SCPs one at a time until you find the one
causing the problem.
(HP Service)*
re-seat the SCP that the channel number(s) points
to, or move the SCP and see if the failure(s) follow
the SCP. If the problems move with the SCP,
replace the SCP.
(HP Service)*
*Must send module to an HP Service Center for repair.
Record information found in FIFO to assist the HP
Service Center in repairing the problem.
Refer to the Command Reference under *TST? for a
list of module functions tested.
Appendix B

Advertisement