Diagnostic:calibration:setup[:Mode]; Diagnostic:calibration:tare[:Otdetect]:Mode; Diagnostic:calibration:tare[:Otdetect]:Mode <Mode - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

DIAGnostic
Comments
Usage

DIAGnostic:CALibration:SETup[:MODE]?

Comments

DIAGnostic:CALibration:TARE[:OTDetect]:MODE

Parameters
Comments
196 HP E1415 Command Reference
When <mode> is set to 1 (the *RST Default) channels are calibrated using the
Least Squares Fit method to provide the minimum error overall (over the entire
output range). When <mode> is 0, channels are calibrated to provide the
minimum error at their zero point. See your SCPs User's Manual for its
accuracy specifications using each mode.
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
*RST Condition: DIAG:CAL:SET:MODE 1
set analog DAC SCP cal mode for best zero accuracy
DIAG:CAL:SET:MODE 0
*CAL?

DIAGnostic:CALibration:SETup[:MODE]?

mode for analog output DAC SCPs.
Returns a 1 when channels are calibrated using the Least Squares Fit method to
provide the minimum error overall (over the entire output range). Returns a 0
when channels are calibrated to provide the minimum error at their zero point.
See your SCPs User's Manual for its accuracy specifications using each mode.
The C-SCPI type is int16.
Related Commands: DIAG:CAL:SET:MOD, *CAL?, CAL:SET
*RST Condition: DIAG:CAL:SET:MODE 1
DIAGnostic:CALibration:TARE[:OTDetect]:MODE < mode >
Transducer Detect current will be turned off or left on (the default mode) during the
CAL:TARE operation.
Parameter
Parameter
Name
Type
mode
boolean (uint 16)
When <mode> is set to 0 (the *RST Default), channels are tare calibrated with
their OTD current off. When <mode> is 1, channels that have their OTD
current on (DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with
their OTD current left on.
By default (*RST) the CALibration:TARE? command will calibrate all
channels with the OTD circuitry disabled. This is done for two reasons: first,
most users do not leave OTD enabled while taking readings, and second, the
CALibration:TARE? operation takes much longer with OTD enabled.
set mode for best zero cal
start channel calibration
returns the currently set calibration
sets whether Open
Range of
Values
0 | 1
Default
Units
volts
Chapter 6