Usage
[SENSe:]TOTalize:RESet:MODE
Parameters
Comments
Usage
[SENSe:]TOTalize:RESet:MODE?
Parameters
Comments
Chapter 6
flt32.
•
<channel> must specify a single channel only.
•
Related Commands: STRAIN:UNST
STRAIN:UNST? (@107)
enter statement here
[SENSe:]TOTalize:RESet:MODE < select >,< ch_list >
totalizer channels in <ch_list>.
Parameter
Parameter
Name
Type
select
discrete (string)
ch_list
string
•
In the INIT mode the total is reset only when the INITiate command is
executed. In the TRIGger mode the total is reset every time a new scan is
triggered.
•
If the channels specified are not on a Frequency/Totalize SCP, an error will be
generated.
•
Related Commands: SENS:FUNC:TOT, INPUT:POLARITY
•
*RST Condition: SENS:TOT:RESET:MODE INIT
SENS:TOT:RESET:MODE TRIG,(@134)
[SENSe:]TOTalize:RESet:MODE? < channel >
totalizer channel in <channel>.
Parameter
Parameter
Name
Type
channel
string
•
Channel must specify a single channel.
query unstrained voltage for channel 7
returns the unstrained voltage set by
STR:UNST
sets the mode for resetting
Range of
Values
INIT | TRIGger
100 - 163
totalizer at channel 34 resets at each
trigger event
returns the reset mode for the
Range of
Values
100 - 163
HP E1415 Command Reference
[SENSe]
Default
Units
seconds
none
Default
Units
none
259