Diagnostic:custom:piecewise; Diagnostic:custom:piecewise <Table_Range>,<Table_Block>, (@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

DIAGnostic
Parameters
Comments
Usage

DIAGnostic:CUSTom:PIECewise

Parameters
Comments
198 HP E1415 Command Reference
Parameter
Parameter
Name
Type
table_range
numeric (float32)
table_block
definite length block
data
ch_list
channel list (string)
<table_block> is a block of 8 bytes that define 4, 16-bit values. SCPI requires
that <table_block> include the definite length block data header. C-SCPI adds
the header for you.
<table_range> specifies the range of voltage that the table covers (from
-<table_range> to +<table_range>). The value you specify must be within 5%
of one of the nominal values from the table above.
<ch_list> specifies which channels may use this custom EU table
Related Commands: [SENSe:]FUNCtion:CUSTom
*RST Condition: All custom EU tables erased
program puts table constants into array table_block
DIAG:CUST:LIN table_block,(@116:123)
SENS:FUNC:CUST:LIN 1,1,(@116:123)
INITiate then TRIGger module
DIAGnostic:CUSTom:PIECewise < table_range >,< table_block >, (@< ch_list >)
downloads a custom piece wise Engineering Unit Conversion table (in
<table_block>) to the HP E1415. Contact your Hewlett-Packard System Engineer
for more information on Custom Engineering Unit Conversion for your application.
Parameter
Parameter
Name
Type
table_range
numeric (float32)
table_block
definite length block
data
ch_list
channel list (string)
<table_block> is a block of 1,024 bytes that define 512 16-bit values. SCPI
requires that <table_block> include the definite length block data header.
C-SCPI adds the header for you.
Range of
Values
.015625 | .03125 | .0625 | .125 | .25 | .5 | 1
| 2 | 4 | 8 | 16 | 32 | 64
see comments
100 - 163
send table to HP E1415 for chs 16-23
link custom EU with chs 16-23
Range of
Values
.015625 | .03125 | .0625 | .125 | .25 | .5 |
1 | 2 | 4 | 8 | 16 | 32 | 64
see comments
100 - 163
Default
Units
volts
none
none
Default
Units
volts
none
none