Output:current:amplitude; Output:current:amplitude? (@<Channel>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Comments
Usage

OUTPut:CURRent:AMPLitude?

Parameters
Comments
Usage
Chapter 6
Select 488E-6 (or MAX) for measuring resistances of less than 8000 Ohms.
Select 30E-6 (or MIN) for resistances of 8000 Ohms and above. amplitude
may be specified in µA (ua).
For resistance temperature measurements ([SENSe:]FUNCtion:TEMPerature)
the Current Source SCP must be set as follows:
MAX (488µA)
MIN (30µA)
When *CAL? is executed, the current sources are calibrated on the range
selected at that time.
When Accepted: Not while INITiated
Related Commands: *CAL?, OUTP:CURR:AMPL?
*RST Condition: MIN
OUTP:CURR:AMPL 488ua,(@116:123)
OUTP:CURR:AMPL 30E-6,(@105)
OUTPut:CURRent:AMPLitude? (@< channel >)
Current Source SCP channel specified by channel.
Parameter
Parameter
Name
Type
channel
channel list (string)
channel must specify a single channel only.
If channel specifies an SCP which is not a Current Source, a +3007, "Invalid
signal conditioning plug-on" error is generated.
Returned Value: Numeric value of amplitude set. The C-SCPI type is float32.
Related Commands: OUTP:CURR:AMPL
OUTP:CURR:AMPLITUDE? (@163)
for RTD,85 | 92 and THER,2250
for THER,5000 | 10000
Set Current Source SCP at channels 16
through 23 to 488 µA
Set Current Source SCP at channel 5 to
30 µA
returns the range setting of the
Range of
Values
100 - 163
Check SCP current set for channel 63
(returns +3.0E-5 or +4.88E-4)
HP E1415 Command Reference
OUTPut
Default
Units
none
221