DIAGnostic:OTDetect[:STATe] < enable >,(@< ch_list >)
HP E1415's "Open Transducer Detection" capability (OTD). When Open Transducer
Detection is enabled, a very high impedance path connects all SCP channels to a
voltage source greater than 16 volts. If an enabled channel has an open transducer,
the input signal becomes the source voltage and the channel returns an input
over-range value. The value returned is +9.91E+37 (ASCII).
Parameter
Parameter
Name
enable
boolean (uint16)
ch_list
channel list (string)
Open Transducer Detection is enabled/disabled on a whole Signal
Conditioning Plug-on basis. Selecting any channel on an SCP selects all
channels on that SCP (8 channels per SCP).
The DIAG:CAL:TARE:MODE <mode> command affects how OTD is
controlled during the CAL:TARE? operation. When <mode> is set to 0 (the
*RST Default), channels are tare calibrated with their OTD current off. When
<mode> is 1, channels that have their OTD current on
(DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with their OTD
current left on.
Related Commands: DIAG:OTDETECT:STATE?,
DIAG:CAL:TARE:MODE
Note
*RST Condition: DIAG:OTDETECT OFF
If OTD is enabled when *CAL?, or CAL:TARE is executed, the module will disable
OTD, wait 1 minute to allow channels to settle, perform the calibration, and then
re-enable OTD.
DIAG:OTD ON,(@100:107,115:123)
DIAG:OTD:STATE ON,(@100,115)
DIAG:OTDETECT:STATE OFF,(@108)
DIAGnostic:OTDetect[:STATe]? (@< channel >)
Transducer Detection" for the SCP containing the specified channel.
Type
1 | 0 | ON | OFF
select OTD for the first and third SCP
(complete channel lists for readability
only)
same function as example above (only
first channel of each SCP specified)
disable OTD for the 8 channels on the
second SCP (only first channel of SCP
specified)
enables and disables the
Range of
Values
100 - 163
returns the current state of "Open
Chapter 6
Default
Units
none
none