Output:type; Output:voltage:amplitude; Output:type? <Channel; Output:voltage:amplitude <Amplitude>,(@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Comments
Usage

OUTPut:TYPE?

Parameters
Comments

OUTPut:VOLTage:AMPLitude

Parameters
Chapter 6
If the channels specified are on an SCP that doesn't support this function an
error will be generated. See your SCP's User's Manual to determine its
capabilities.
PASSive configures the digital channel/bit to be passive (resistor) pull-up to
allow you to wire-or more than one output together.
ACTive configures the digital channel/bit to both source and sink current.
Related Commands: SOURce:PULSe:POLarity, OUTPut:TYPE?
*RST Condition: OUTP:TYPE ACTIVE (for TTL compatibility)
OUTP:TYPE PASS,(@140:143)
OUTPut:TYPE? < channel >
channel.
Parameter
Parameter
Name
Type
channel
string
Channel must specify a single channel.
If the channel specified is not on a digital SCP, an error will be generated.
Returned Value: returns PASS or ACT. The type is string.
*RST Condition: returns ACT
OUTPut:VOLTage:AMPLitude < amplitude >,(@< ch_list >)
voltage on programmable Strain Bridge Completion SCPs pointed to by <ch_list>
(the HP E1511 for example). This command is not used to set output voltage on SCPs
like the HP E1531 Voltage Output SCP.
Parameter
Parameter
Name
Type
amplitude
numeric (float32)
ch_list
channel list (string)
make channels 40 to 43 passive pull-up
returns the output drive characteristic for a digital
Range of
Values
100 - 163
Range of
Values
MIN | 0 |1 | 2 | 5 | 10 | MAX
100 - 163
HP E1415 Command Reference
OUTPut
Default
Units
none
sets the excitation
Default
Units
none
none
227