Source:function[:Shape]:Square; Source:pulm[:State]; Source:function[:Shape]:Square (@<Ch_List>); Source:pulm[:State] <Enable>,(@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Parameters
Comments

SOURce:FUNCtion[:SHAPe]:SQUare

Parameters
Comments

SOURce:PULM[:STATe]

Parameters
Comments
Chapter 6
Parameter
Parameter
Name
Type
ch_list
string
This PULSe channel function is further defined by the SOURce:FM:STATe
and SOURce:PULM:STATe commands. If the FM state is enabled then the
frequency modulated mode is active. If the PULM state is enabled then the
pulse width modulated mode is active. If both the FM and the PULM states are
disabled then the PULSe channel is in the single pulse mode.
SOURce:FUNCtion[:SHAPe]:SQUare (@< ch_list >)
output a square wave (50% duty cycle) on the channels in <ch_list>.
Parameter
Parameter
Name
Type
ch_list
string
The frequency control for these channels is provided by the algorithm
language function:.
O143 = 2000 /* set channel 43 to 2KHz */
SOURce:PULM[:STATe] < enable >,(@< ch_list >)
modulated mode for the PULSe channels in <ch_list>.
Parameter
Parameter
Name
Type
enable
boolean (uint16)
ch_list
string
This command is coupled with the SOURce:FM command. If the FM state is
enabled then the PULM state is disabled. If the PULM state is enabled then the
FM state is disabled. If both the FM and the PULM states are disabled then the
PULSe channel is in the single pulse mode.
If the channels specified are not on a Frequency/Totalize SCP, an error will be
generated.
Range of
Values
100 - 163
sets the SOURce function to
Range of
Values
100 - 163
enable the pulse width
Range of
Values
1 | 0 | ON | OFF
100 - 163
HP E1415 Command Reference
SOURce
Default
Units
none
Default
Units
none
Default
Units
none
none
263