Command
Sequence
CALibration:VALue:VOLTage
Parameters
Comments
Command
Sequence
Chapter 6
multimeter connected to the H Cal, L Cal, H ohm, and L ohm terminals on
the Terminal Module, or the V H, V L, Ω H, and Ω L terminals on the Cal Bus
connector.
•
Use the CAL:CONF:RES command to configure the reference resistor for
measurement at the Calibration Bus connector.
•
When Accepted: Not while INITiated
•
Related Commands: CAL:CONF:RES, CAL:STORE ADC
CAL:CONF:RES
( now measure ref resistor with external DMM )
CAL:VAL:RES <measured value>
CALibration:VALue:VOLTage < ref_volts >
DC reference source to the module for A/D calibration.
Parameter
Parameter
Name
Type
ref_volts
numeric (float32)
•
The value sent must be for the currently configured range and output (zero or
full scale) as set by the previous CAL:CONF:VOLT <range>, ZERO |
FSCale command. Full scale values must be within 10% of .0625, .25, 1, 4, or
10 (the voltage reference provides 10VDC on the 16V range).
•
ref_volts may be specified in millivolts (mv).
•
A measurement of the source voltage can be made with an external multimeter
connected to the H Cal, and L Cal terminals on the Terminal Module, or the
V H, and V L terminals on the Cal Bus connector.
•
Use the CAL:CONF:VOLT command to configure the on-board voltage
source for measurement at the Calibration Bus connector.
•
When Accepted: Not while INITiated
•
Related Commands: CAL:CONF:VOLT, CAL:STORE ADC
CAL:CONF:VOLTAGE 4,FSCALE
*OPC?
enter statement
(now measure voltage with external DMM)
Send measured value to module
sends the value of the just-measured
Range of
Values
must be within 10% of range nominal
Wait for operation to complete
HP E1415 Command Reference
CALibration
Default
Units
volts
193