[Sense:]Function:strain:fbending; [Sense:]Function:strain:fbpoisson; [Sense:]Function:strain:fpoisson; [Sense:]Function:strain:hbending - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

[SENSe]

[SENSe:]FUNCtion:STRain:FBENding

[SENSe:]FUNCtion:STRain:FBPoisson

[SENSe:]FUNCtion:STRain:FPOisson

[SENSe:]FUNCtion:STRain:HBENding

[SENSe:]FUNCtion:STRain:HPOisson

[SENSe:]FUNCtion:STRain[:QUARter]

Note
248 HP E1415 Command Reference
[SENSe:]FUNCtion:STRain:FBENding [< range >,](@< ch_list >)
[SENSe:]FUNCtion:STRain:FBPoisson [< range >,](@< ch_list >)
[SENSe:]FUNCtion:STRain:FPOisson [< range >,](@< ch_list >)
[SENSe:]FUNCtion:STRain:HBENding [< range >,](@< ch_list >)
[SENSe:]FUNCtion:STRain:HPOisson [< range >,](@< ch_list >)
[SENSe:]FUNCtion:STRain[:QUARter] [< range >,](@< ch_list >)
Note on Syntax: Although the strain function is comprised of six separate SCPI
commands, the only difference between them is the bridge type they specify to the
strain EU conversion algorithm.
[SENSe:]FUNCtion:STRain:<bridge_type> [<range>,](@<ch_list>)
strain EU conversion with the channels specified by ch_list to measure the bridge
voltage. See "Linking Input Channels to EU Conversion" on page 64.
The following table relates the command syntax to bridge type. See your Strain SCP
user's manual for bridge schematics and field wiring information.
Command
:FBENding
:FBPoisson
:FPOisson
:HBENding
:HPOisson
[:QUARter]
Because of the number of possible strain gage configurations, the driver must
generate any Strain EU conversion tables and download them to the instrument when
INITiate is executed. This can cause the time to complete the INIT command to
exceed 1 minute.
Bridge Type
Full Bending Bridge
Full Bending Poisson Bridge
Full Poisson Bridge
Half Bending Bridge
Half Poisson Bridge
Quarter Bridge (default)