SAMPle
Subsystem Syntax
SAMPle:TIMer
Parameters
Comments
Usage
Chapter 6
The SAMPle subsystem provides commands to set and query the interval between
channel measurements (pacing).
SAMPle
:TIMer <interval>
:TIMer?
SAMPle:TIMer < interval >
is used to provide additional channel settling time. See "Settling Characteristics" on
page 110.
Parameter
Parameter
Name
Type
interval
numeric (float32)
(string)
The minimum interval is 10 µ seconds. The resolution for interval is 2.5
•
µsecond.
•
If the Sample Timer interval multiplied by the number of channels in the
specified Scan List is longer than the Trigger Timer interval, at run time a
"Trigger too fast" error will be generated.
•
the SAMP:TIMER interval can change the effect of the
SENS:CHAN:SETTLING command. ALG:CHAN:SETT specifies the
number of times a channel measurement should be repeated. The total settling
time per channel then is (SAMP:TIMER <interval>) X (<chan_repeats> from
SENS:CHAN:SETT)
•
When Accepted: Not while INITiated
•
Related Commands: SENSE:CHAN:SETTLING, SAMP:TIMER?
•
*RST Condition: Sample Timer for all Channel Lists set to 1.0E-5 seconds.
SAMPLE:TIMER 50E-6
sets the time interval between channel measurements. It
Range of
1.0E-5 to 16.3825E-3 |
MIN | MAX
Pace measurements at 50µsecond
intervals
HP E1415 Command Reference
Default
Values
seconds
Units
231