Calibration:configure:resistance - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

CALibration
Subsystem Syntax

CALibration:CONFigure:RESistance

Comments
186 HP E1415 Command Reference
Figure 6-3. Levels of Working Calibration
CALibration
:CONFigure
:RESistance
:VOLTage < range >, ZERO | FS
:SETup
:SETup?
:STORe ADC | TARE
:TARE (@< ch_list >)
:RESet
:TARE?
:VALue
:RESistance < ref_ohms >
:VOLTage < ref_volts >
:ZERO?

CALibration:CONFigure:RESistance

the Calibration Bus. A four-wire measurement of the resistor can be made with an
external multimeter connected to the H Cal, L Cal, H ohm, and L ohm terminals on
the Terminal Module, or the V H, V L, Ω H, and Ω L terminals on the Cal Bus
connector.
Related Commands: CAL:VAL:RES, CAL:STOR ADC
When Accepted: Not while INITiated
connects the on-board reference resistor to
Chapter 6