[SENSe]
Usage
[SENSe:]REFerence:CHANnels
Parameters
Comments
Usage
[SENSe:]REFerence:TEMPerature
254 HP E1415 Command Reference
again.
•
Related Commands: SENSE:FUNC:TEMP
*RST Condition: Reference temperature is 0 °C
•
sense the reference temperature on channel 20 using an RTD
SENSE:REF RTD,92,(@120)
[SENSe:]REFerence:CHANnels (@< ref_channel >),(@< ch_list >)
specified by <ref_channel> to appear in the scan list just before the channel(s)
specified by <ch_list>. This command is used to include the thermocouple reference
temperature channel in the scan list before other thermocouple channels are
measured.
Parameter
Parameter
Name
Type
ref_channel
channel list (string)
ch_list
channel list (string)
•
Use SENS:FUNC:TEMP to configure channels to measure thermocouples.
Then use SENS:REF to configure one or more channels to measure an
isothermal reference temperature. Now use SENS:REF:CHAN to group the
reference channel with its thermocouple measurement channels in the scan list.
•
If thermocouple measurements are made through more than one isothermal
reference panel, you will set up a reference channel for each. Execute the
SENS:REF:CHAN command for each reference/measurement channel group.
•
Related commands: SENS:FUNC:TEMP, SENS:REF
•
*RST Condition: Scan List contains no channel references.
SENS:FUNC:TEMP TC,E,.0625,(@108:115) E type TCs on channels 8 through 15
SENS:REF THER,5000,1,(@106)
SENS:REF RTD,85,.25,(@107)
SENS:REF:CHAN (@106),(@108:111)
SENS:REF:CHAN (@107),(@112:115)
[SENSe:]REFerence:TEMPerature < degrees_c >
temperature in the Reference Temperature Register. Use when the thermocouple
reference junction is kept at a controlled temperature.
causes channel
Range of
Values
100 - 163
100 - 163
Reference ch is thermistor at channel 6
Reference ch is RTD at channel 7
Thermistor measured before chs 8 - 11
RTD measured before chs 12 - 15
stores a fixed reference junction
Default
Units
none
none
Chapter 6