Usage
[SENSe:]FUNCtion:TOTalize
Parameters
Comments
Usage
[SENSe:]FUNCtion:VOLTage[:DC]
Chapter 6
Source SCP and Sense Amplifier SCP setup at the time of execution. If SCP
settings are changed, those channels are no longer calibrated. *CAL? must be
executed again.
•
See "Linking Input Channels to EU Conversion" on page 64.
•
When Accepted: Not while INITiated
•
Related Commands: *CAL?, OUTP:CURR (for RTDs and Thermistors),
SENS:REF, and SENS:REF:TEMP (for Thermocouples)
•
*RST Condition: SENSE:FUNC:VOLT AUTO,(@100:163)
Link two channels to the K type thermocouple temperature conversion
SENS:FUNC:TEMP TCOUPLE,K,(@101,102)
Link channel 0 to measure reference temperature using 5K thermistor
SENS:REF THER,5000,(@100)
[SENSe:]FUNCtion:TOTalize < ch_list >
channels in <ch_list>.
Parameter
Parameter
Name
Type
ch_list
string
•
The totalize function counts rising edges of digital transitions at
Frequency/Totalize SCP channels. The counter is 24 bits wide and can count
up to 16,777,215.
•
The SENS:TOT:RESET:MODE command controls which events will reset the
counter.
•
If the channels specified are not on a Frequency/Totalize SCP, an error will be
generated.
•
Related Commands: SENS:TOT:RESET:MODE, INPUT:POLARITY
•
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital
SCP channels.
SENS:FUNC:TOT (@134)
[SENSe:]FUNCtion:VOLTage[:DC] [< range >,](@< ch_list >)
sets the SENSe function to TOTalize for
Range of
Values
100 - 163
channel 34 is a totalizer
links the specified
HP E1415 Command Reference
[SENSe]
Default
Units
none
251