Fetch - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

FETCh?

Subsystem Syntax
Comments
204 HP E1415 Command Reference

FETCh?

returns readings stored in VME memory.
This command is only available in systems using an HP E1405B or
HP E1406A command module.
FETCH? does not alter the readings stored in VME memory. Only the *RST or
INIT... commands will clear the readings in VME memory.
The format of readings returned is set using the FORMat[:DATA] command.
Returned Value: REAL,32, REAL,64, and PACK,64, readings are returned in
the IEEE-488.2-1987 Definite Length Arbitrary Block Data format. This data
return format is explained in "Arbitrary Block Program and Response Data"
on page 160. For REAL,32, readings are 4 bytes in length. For REAL 64, and
PACK, 64, readings are 8 bytes in length.
PACKed,64 returns the same values as REAL,64 except for Not-a-Number
(NaN), IEEE +INF and IEEE -INF. The NaN, IEEE +INF and IEEE -INF
values returned by PACKed,64 are in a form compatible with HP Workstation
BASIC and HP BASIC/UX. Refer to the FORMat command for the actual
values for NaN, +INF, and -INF.
ASCii is the default format.
ASCII readings are returned in the form ±1.234567E±123. For example 13.325
volts would be +1.3325000E+001. Each reading is followed by a comma (,). A
line feed (LF) and End-Or-Identify (EOI) follow the last reading.
Related Commands: MEMory Subsystem, FORMat[:DATA]
*RST Condition: MEMORY:VME:ADDRESS 240000;
MEMORY:VME:STATE OFF; MEMORY:VME:SIZE 0