Input:filter[:Lpass]:Frequency; Input:filter[:Lpass][:State]; Input:filter[:Lpass]:Frequency? (@<Channel>); Input:filter[:Lpass][:State] <Enable>,(@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Usage

INPut:FILTer[:LPASs]:FREQuency?

Parameters
Comments
Usage

INPut:FILTer[:LPASs][:STATe]

Parameters
Chapter 6
*RST Condition: set to MIN
INP:FILT:FREQ 100,(@100:119)
INPUT:FILTER:FREQ 2,(@155)
INPut:FILTer[:LPASs]:FREQuency? (@< channel >)
currently set for channel. Non-programmable SCP channels may be queried to
determine their fixed cutoff frequency. If the channel is not on an input SCP, the
query will return zero.
Parameter
Parameter
Name
Type
channel
channel list (string)
channel must specify a single channel only.
This command is for programmable filter SCPs only.
Returned Value: Numeric value of Hz as set by the INP:FILT:FREQ
command. The C-SCPI type is float32.
When Accepted: Not while INITiated
Related Commands: INP:FILT:LPAS:FREQ, INP:FILT:STATE
*RST Condition: MIN
INPUT:FILTER:LPASS:FREQUENCY? (@155)
INP:FILT:FREQ? (@100)
INPut:FILTer[:LPASs][:STATe] < enable >,(@< ch_list >)
programmable filter SCP channel. When disabled (enable=OFF), these channels are
in their "pass through" mode and provide no filtering. When re-enabled (enable=ON),
the SCP channel reverts to its previously programmed setting.
Parameter
Parameter
Name
Type
enable
boolean (uint16)
ch_list
channel list (string)
Set cutoff frequency of 100 Hz for first 20
channels
Set cutoff frequency of 2 Hz for channel
55
returns the cutoff frequency
Range of
Values
100 - 163
Check cutoff freq on channel 55
Check cutoff freq on channel 0
enables or disables a
Range of
Values
1 | 0 | ON | OFF
100 - 163
HP E1415 Command Reference
INPut
Default
Units
none
Default
Units
none
none
211