[Sense:]Strain:excitation; [Sense:]Strain:excitation <Excite_V>,(@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Note
Parameters
Comments
Usage

[SENSe:]STRain:EXCitation

Parameters
Comments
Usage
Chapter 6
This reference temperature is used to compensate all subsequent thermocouple
measurements until the register is overwritten by another SENSE:REF:TEMP value
or by scanning a channel linked with the SENSE:REFERENCE command. If used,
SENS:REF:TEMP must be executed before scanning any thermocouple channels.
Parameter
Parameter
Name
Type
degrees_c
numeric (float32)
This command is used to specify to the HP E1415 the temperature of a
controlled temperature thermocouple reference junction.
When Accepted: Not while INITiated
Related Commands: FUNC:TEMP TC...
*RST Condition: Reference temperature is 0 °C
SENSE:REF:TEMP 40
[SENSe:]STRain:EXCitation < excite_v >,(@< ch_list >)
voltage value to be used to convert strain bridge readings for the channels specified
by <ch_list>. This command does not control the output voltage of any source.
Parameter
Parameter
Name
Type
excite_v
numeric (flt32)
ch_list
channel list (string)
<ch_list> must specify the channel used to sense the bridge voltage, not the
channel position on a Bridge Completion SCP.
Related Commands: SENSE:STRAIN:..., SENSE:FUNC:STRAIN...
*RST Condition: 3.9V
STRAIN:EXC 4,(@100:107)
Range of
Values
-126 to +126
subsequent thermocouple conversion will
assume compensation junction at
40 degrees C
specifies the excitation
Range of
Values
.01 - 99
100 - 163
set excitation voltage for channels 0
through 7
HP E1415 Command Reference
[SENSe]
Default
Units
none
Default
Units
volts
none
255