HP VXI 75000 C Series User's And Scpi Programming Manual page 253

Algorithmic closed loop controller
Table of Contents

Advertisement

Parameters
Comments
Chapter 6
is used to compensate all subsequent thermocouple measurements until the register
is overwritten by another reference measurement or by specifying a constant
reference temperature with the SENSE:REF:TEMP command. If used, the reference
junction channel must be scanned before any thermocouple channels. Use the
SENSE:REF:CHANNELS command to place the reference measuring channel into
the scan list ahead of the thermocouple measuring channels.
Parameter
Parameter
Name
Type
type
discrete (string)
sub_type
numeric (float32)
numeric (float32)
range
numeric (float32)
ch_list
channel list (string)
See "Linking Input Channels to EU Conversion" on page 64.
The <range> parameter: The HP E1415 has five ranges: .0625VDC, .25VDC,
1VDC, 4VDC, and 16VDC. To select a range, simply specify the range value
(for example, 4 selects the 4VDC range). If you specify a value larger than
one of the first four ranges, the HP E1415 selects the next higher range (for
example, 4.1 selects the 16VDC range). Specifying a value larger than 16
causes an error. Specifying 0 selects the lowest range (.0625VDC).
Specifying AUTO selects auto range. The default range (no range parameter
specified) is auto range.
If you are using amplifier SCPs, you should set them first and keep their
settings in mind when specifying a range setting. For instance, if your expected
signal voltage is to be approximately .1VDC and the amplifier SCP for that
channel has a gain of 8, you must set <range> no lower than 1VDC or an input
out-of-range condition will exist.
The <type> parameter specifies the sensor type that will be used to determine
the temperature of the isothermal reference panel. <type> CUSTom is
pre-defined as Type E with 0°C reference junction temp and is not
re-defineable.
For <type> THERmistor, the <sub_type> parameter may be specified in ohms
or kohm.
The *CAL? command calibrates resistance channels based on Current Source
SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings
are changed, those channels are no longer calibrated. *CAL? must be executed
Range of
Values
THERmistor | RTD | CUSTom
for THER use 5000
for RTD use 85 | 92
for CUSTom use 1
see comments
100 - 163
HP E1415 Command Reference
[SENSe]
Default
Units
none
Ohm
none
none
VDC
none
253