Usage
DIAGnostic:CALibration:TARE[:OTDetect]:MODE?
Comments
DIAGnostic:CHECksum?
Comments
Usage
DIAGnostic:CUSTom:LINear
Chapter 6
However, for users who intend to take readings with OTD enabled, setting
DIAG:CAL:TARE:OTD:MODE to 1, will force the CAL:TARE? command to
perform calibration with OTD enabled on channels so specified by the user
with the DIAG:OTD command.
•
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
•
*RST Condition: DIAG:CAL:TARE:MODE 0
configure OTD on during CAL:TARE
DIAG:CAL:TARE:MODE 1
CAL:TARE?
DIAGnostic:CALibration:TARE[:OTDetect]:MODE?
mode for controlling Open Transducer Detect current while performing CAL:TARE?
operation.
•
Returns a 0 when OTD current will be turned off during CAL:TARE?. Returns
1 when OTD current will be left on during CAL:TARE? operation. The
C-SCPI type is int16.
•
Related Commands: DIAG:CAL:TARE:MOD, DIAG:OTD, CAL:TARE?
•
*RST Condition: DIAG:CAL:TARE:MODE 0
DIAGnostic:CHECksum?
returned value of 1 indicates that Flash memory contents are correct. A returned value
of 0 indicates that the Flash Memory is corrupted, or has been erased.
•
Returned Value: Returns 1 or 0. The C-SCPI type is int16.
DIAG:CHEC?
DIAGnostic:CUSTom:LINear < table_range >,< table_block >, (@< ch_list >)
downloads a custom linear Engineering Unit Conversion table (in <table_block>) to
the HP E1415. Contact your Hewlett-Packard System Engineer for more information
on Custom Engineering Unit Conversion for your application.
set mode for OTD to stay on
start channel tare cal.
performs a checksum operation on Flash Memory. A
Checksum Flash Memory, return 1 for
OK, 0 for corrupted
HP E1415 Command Reference
DIAGnostic
returns the currently set
197