Input:gain; Input:low; Input:gain? (@<Channel>); Input:low <Wvolt_Type>,(@<Ch_List>) - HP VXI 75000 C Series User's And Scpi Programming Manual

Algorithmic closed loop controller
Table of Contents

Advertisement

Usage

INPut:GAIN?

Parameters
Comments
Usage

INPut:LOW

Chapter 6
Sending MIN will program the lowest gain.
Sending a value for gain that is greater than the highest or less than the lowest
setting allowable for the SCP will generate a -222 "Data out of range" error.
When Accepted: Not while INITiated
Related Commands: INP:GAIN?
*RST Condition: gain set to MIN
INP:GAIN 8,(@100:119)
INPUT:GAIN 64,(@155)
INPut:GAIN? (@< channel >)
channel is not on an input SCP, the query will return zero.
Parameter
Parameter
Name
Type
channel
channel list (string)
channel must specify a single channel only.
If the channel specified does not have a programmable amplifier, INP:GAIN?
will return the nominal as-designed gain for that channel.
Returned Value: Numeric value as set by the INP:GAIN command. The
C-SCPI type is float32.
When Accepted: Not while INITiated
Related Commands: INP:GAIN
*RST Condition: gain set to 1
INPUT:GAIN? (@105)
INP:GAIN? (@100)
INPut:LOW < wvolt_type >,(@< ch_list >)
Strain Bridge SCP channel specified by <ch_list>. LO can be connected to the
Wagner Voltage ground or left floating.
Set gain of 8 for first 20 channels
Set gain of 64 for channel 55
returns the gain currently set for channel. If the
Range of
Values
100 - 163
Check gain on channel 5
Check gain on channel 0
controls the connection of input LO at a
HP E1415 Command Reference
INPut
Default
Units
none
213