[SENSe]
[SENSe:]FREQuency:APERture?
Parameters
Comments
[SENSe:]FUNCtion:CONDition
Parameters
Comments
Usage
242 HP E1415 Command Reference
[SENSe:]FREQuency:APERture? < channel >
time.
Parameter
Parameter
Name
Type
channel
string
•
If the channel specified is on an SCP that doesn't support this function, an error
will be generated. See your SCP's User's Manual for its capabilities.
•
Related Commands: SENSe:FREQuency:APERture
•
Returned Value: returns numeric gate time in seconds, The type is float32.
[SENSe:]FUNCtion:CONDition < ch_list >
digital state for channels in <ch_list>. Also configures digital SCP channels as inputs
(this is the *RST condition for all digital I/O channels).
Parameter
Parameter
Name
Type
ch_list
string
•
The HP E1533 SCP senses 8 digital bits on each channel specified by this
command. The HP E1534 SCP senses 1 digital bit on each channel specified
by this command.
•
If the channels specified are not on a digital SCP, an error will be generated.
•
Use the INPut:POLarity command to set input logical sense.
•
Related Commands: INPut:POLarity
•
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital
SCP channels.
To set second 8-bits of HP E1533 at SCP position 4, and upper 4-bits of HP E1534
at SCP position 5 to digital inputs send:
SENS:FUNC:COND (@133,144:147)
returns the frequency counting gate
Range of
Values
100 - 163
sets the SENSe function to input the
Range of
Values
100 - 163
Default
Units
none
Default
Units
none
Chapter 6