Analog Output Offset Adjustment - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
Press SET> to select the analog output channel to be configured:
DISPLAYED AS =
CONC_OUT_1 =
CONC_OUT_2 =
CONC_OUT_3 =
TEST OUTPUT =
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET>
EDIT
These keys increase / decrease the analog
output by 100, 10 or 1 counts. Continue
adjustments until the voltage measured at the
output of the analyzer and/or the input of the
recording device matches the value in the upper
right hand corner of the display to the tolerance
The concentration display will not change. Only
the voltage reading of your voltmeter will change.

6.7.3.4. Analog Output Offset Adjustment

Some analog signal recorders require that the zero signal is significantly different from the
baseline of the recorder in order to record slightly negative readings from noise around the
zero point. This can be achieved in the M200E by defining a zero offset, a small voltage
(e.g., 10% of span), which can be added to the signal of individual output channels by
pressing SETUP-MORE-DIAG-ENTR-NEXT-NEXT and the following keys:
044100102 Rev A
DIAG AIO
ENTR
EXIT
DIAG AIO
< SET
EXIT
DIAG AIO
CHANNEL
A1
A2
A3
DIAG AIO
A4
< SET
EXIT
listed in Table 6-10.
CONC_OUT_1 RANGE: 5V
SET>
EDIT
CONC_OUT_1 REC OFS: 0 mV
SET>
EDIT
CONC_OUT_1 AUTO CAL: OFF
< SET
SET>
EDIT
CONC_OUT_2 CALIBRATED: NO
CAL
DIAG AIO
CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_1 CALIBRATED: YES
< SET
CAL
Operating Instructions
EXIT
EXIT
If AutoCal is ON, go to
Section 6.7.3
EXIT
EXIT
EXIT ignores the
new setting.
ENTR accepts the
new setting.
EXIT
65