10.8.3
Adjustment of the on-site display
The on-site display can be adjusted in the following menu:
Setup ® Advanced setup ® Display (® ä 161)
10.9
Configuration management
After commissioning, you can save the current device configuration, copy it to another measuring
point or restore the previous device configuration. You can do so using the Configuration
management parameter and its options.
Navigation path
Setup ® Advanced setup ® Conf.backup disp ® Config. managem.
Functions of the parameter options
Options
Description
Execute backup
A backup copy of the current device configuration in the HistoROM is saved to the
display module of the device. The backup copy comprises the transmitter data of the
device.
Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device. The backup copy comprises the transmitter data of the device.
Duplicate
The transmitter configuration from another device is duplicated to the device using the
display module.
Compare
The device configuration saved in the display module is compared to the current device
configuration of the HistoROM.
Clear backup data
The backup copy of the device configuration is deleted from the display module of the
device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local display and a
message on the processing status appears on the display.
10.10 Protection of the settings against unauthorized
changes
There are two ways to protect the settings against unauthorized changes:
• Via locking switch (hardware locking) (® ä 66)
• Via parameter settings (software locking) (® ä 67)
Commissioning via operating menu (On-site display, FieldCare)
83
Need help?
Do you have a question about the Levelflex FMP51 and is the answer not in the manual?
Questions and answers