Larson Davis SoundAdvisor 831C Reference Manual page 121

Sound level meter
Hide thumbs Also See for SoundAdvisor 831C:
Table of Contents

Advertisement

FIGURE 15-2
Continuous Run Mode
SoundAdvisor Model 831C
d.
If desired, select Interval Time Sync. This forces a new measurement record to
begin every n minutes (or hours), where n is the value of Interval Time.
If you enter an Interval Time (hh:mm) value less than 0:01, the measure-
ment interval syncs to the 1 min interval boundary.
If you enter any other Interval Time (hh:mm) value of 01:00 or more, the
measurement syncs to the hour.
e.
If you want to send a measurement interval exceedance alert, select a checkbox
in-line with Alerts.
f.
Select a Trigger Source and Trigger Level (dB) for the alert.
1.
93
g.
Select any additional settings on the Setup Manager tabs. For more
information, see 6.2.2 Customizing Setup Values On the Setup Manager Tabs.
h.
Press Close (
i.
Set it as the Active setup file.
Selecting Single Block Timer Mode With Measurement History
a.
Enter a Start (y-m-d) date and Time, and an End (y-m-d) date and Time,
where Time is the time of day.
b.
Select the Enable Measurement History checkbox.
c.
If you want to send a measurement interval exceedance alert, select your
preferred method in-line with Alerts, and select a Trigger Source and Trigger
Level (dB) for the alert. The Trigger Level cannot be less than 0.1 (dB).
Trigger Sources for this alert include the following statistics: LAeq, LCeq,
TAKE NOTE
LZeq, Lmax, Lmin, and all 6 values of Ln. If the current dB level is greater than the
Trigger Level (dB) of the selected Trigger Source at any time during the current
measurement interval, the 831C sends an alert notification via email or text when
the interval ends.
1.
Trigger Level cannot be <0.1 (dB)
Trigger Sources for this alert include the following
TAKE NOTE
statistics: LAeq, LCeq, LZeq, Lmax, Lmin, and all 6 values of Ln.
If the current dB level is greater than the Trigger Level (dB) of the
selected Trigger Source at any time during the current
measurement interval, the 831C sends an alert notification via
email or text when the interval ends.
/*/
) and save your Setup.
Setting Up a Measurement With Measurement History