Calibration description and purpose
Normalization (TaqMan
• The software extracts a spectral profile for each normalization standard.
• The software uses the normalization standards to obtain a correction
factor for the specific dyes, for example, FAM
• The correction factor is used to generate comparable component signal
levels across instruments.
View the calibration status and set reminders
View the calibration status on the instrument
The Status and Reminders screen also allows you to perform the following functions:
• Transfer the calibration results (see
• Transfer the calibration data files (see
• View the calibration images (see
1. In the home screen, tap
reminders.
The Status and Reminders screen is displayed. It shows the following information:
• Type of calibration
Note:
·
For dye calibration, the type of calibration is listed as the dye that was calibrated, for
example, FAM.
·
Normalization calibration is required for a TaqMan
is listed as FAM/ROX or VIC/ROX.
• Status of the calibration
• Last calibration
• Date that the calibration expires
• If a reminder is set
2. Tap a calibration row, then tap View to view the history of that specific calibration type.
• Date performed
• Result of the calibration
3. Tap Cancel to return to the Status and Reminders screen.
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
Array Card block only)
page
page
(Settings)4Maintenance and service4Calibration4Status and
Chapter 7 Calibrate and verify instrument performance
View the calibration status and set reminders
™
dye or VIC
™
dye.
94)
page
94)
93)
™
Array Card block. The type of calibration
• Whether the calibration is current
• Operator
™
7 Pro Real-Time PCR System User Guide
Pass criteria
Normalization standards have
sufficient signals to generate a
correction factor.
7
87