Calibration description and purpose
Dye
• The software extracts a spectral profile for each dye standard, then
produces a set of spectral profiles plotted as fluorescence vs filter.
• The software uses calibration data to characterize and distinguish the
individual contribution of each dye in the total fluorescence signals
collected by the instrument.

View the calibration status and set reminders

View calibration
status and set
reminders in the
instrument
View calibration
status and set
reminders in the
Cloud
QuantStudio
1 Real-Time PCR System Installation, Use, and Maintenance Guide
1. In the home screen, touch
Service4Calibrations4History and Reminders.
2. In the Calibration Reminders screen, view the status of each calibration type.
3. (Optional) Touch a calibration row to view the history of that specific calibration
type, then touch Done.
4. Touch Edit to set the calibration reminder settings. For each calibration type:
a. Slide the control On to enable the calibration reminder.
b. Edit the Exp interval and Remind me fields.
c. Touch Save.
5. (Optional) To transfer the calibration report, touch Export then follow the
directions on the screen.
6. Touch Done.
Note: The calibration reminders feature requires a connection between the
instrument and a computer network.
1. In the Thermo Fisher Cloud, click
2. Select any of your registered instruments.
3. In the Summary tab:
• Click
Calibrations to view the status of each calibration type.
• (Optional) Click
time table and enter the notification email address(es).
Note: More than one email address can receive the calibration reminders.
Note: The settings are automatically saved.
• (Optional) In the Downloads section, click Maintenance Summary.pdf to
download the calibration status report.
Chapter 5 Calibrate and verify instrument performance

View the calibration status and set reminders

Dye spectra peak within the same
filter as their group.
Settings4Maintenance and
for the Instrument Connect page.
Calibration Reminders to set the calibration reminder
5
Pass Criteria
51