Figure 10 Home screen during a run
Arrow
1
Actions menu
2
Edit a plate file during a run
The targets and samples can be added or edited after a run has started. The method cannot be edited
after a run has started.
1. In the home screen during a run, tap
See Figure 10
2. Tap Set up.
3. In the Plate tab, tap a well or multiple wells.
Note: Use the plate view to edit or assign a well (see "Options to view a plate layout"
4. Tap the Sample Information pane, then tap Samples.
Note: If no wells are selected, the pane is labeled Legend.
5. Enter a sample name, then select a sample type from the dropdown list.
• Unknown
• Standard
• No template control (NTC)
6. Tap Target/SNP, then select Target or Assay from the dropdown list.
7. Tap a row in the table, then tap Close4Done.
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
2
on page
71.
1
(Actions).
™
7 Pro Real-Time PCR System User Guide
Chapter 5 Options for a run
Edit a plate file during a run
on page
5
63).
71