Chapter 4 Edit a plate before starting a run
4
Options to edit the well details of a plate file
4. Tap a row in the table, then delete the information in each field.
• Reagent name
• Reagent type
• Lot number
5. Tap Save reagent.
6. Tap Done.
Add or edit the reagent details for a TaqMan
1. In the Plate tab, tap a port.
See Figure 9
2. Tap the Legend pane, then tap Reagents.
3. Tap a row in the table.
4. Tap Add new, then enter the information.
• Reagent name
• Reagent type
• Lot number
5. Tap Save reagent, then tap Done.
Apply the reagents to a TaqMan
Add the reagents before they are applied to a plate (see "Add or edit the reagent details for a TaqMan
Array Card"
on page
A reagent can only be applied by port. It cannot be applied by individual well.
1. In the Plate tab, tap a port.
See Figure 9
2. Tap the Legend pane, then tap Reagents.
3. Tap a row in the table to apply it to the port.
4. Tap Done.
Edit a passive reference dye
A passive reference dye normalizes the fluorescent reporter signal in real-time PCR.
1. In the Plate tab, tap Actions4Edit passive reference dye.
The Passive Reference Dye screen is displayed.
2. Select the passive reference dye.
68
QuantStudio
on page
65.
™
68).
on page
65.
™
6 Pro Real-Time PCR System and QuantStudio
• SKU (Part number)
• Expiration date
• Quencher dye
™
Array Card
• SKU (Part number)
• Expiration date
Array Card
™
7 Pro Real-Time PCR System User Guide
™
Need help?