Chapter 7 Calibrate And Verify Instrument Performance; Instrument And Block Calibration - Thermo Scientific Applied Biosystems QuantStudio 6 Pro User Manual

Real-time pcr system
Table of Contents

Advertisement

7
Instrument and block calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Calibration and verification schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Calibration descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
View the calibration status and set reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Export a calibration report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
TaqMan
Array Card block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Perform verification using RNase P plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Calibrate custom dyes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Calibrate for a custom melt curve run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120

Instrument and block calibration

The QuantStudio
calibration status applies to the combination of the instrument and the block.
If a block is moved to an instrument that it has not been calibrated on, that combination of block and
instrument needs to be calibrated. This applies even if the block has previously been calibrated on a
different instrument.
If a block is moved back to an instrument that is has previously been calibrated on, it does not need to
be calibrated again. It needs to be calibrated again if the calibration is expired.
A run can still be started if a calibration of the block on the instrument is expired. A run cannot be
started if a block has never been calibrated on the instrument.
84
QuantStudio
Calibrate and verify instrument
6 Pro System and the QuantStudio
6 Pro Real-Time PCR System and QuantStudio
performance
Array Card . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
7 Pro System have interchangeable blocks. The
7 Pro Real-Time PCR System User Guide

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Applied Biosystems QuantStudio 6 Pro and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Applied biosystems quantstudio 7 pro

Table of Contents