Chapter 11 Configure the instrument settings
11
Configure maintenance and service settings
The Hardware Test screen is displayed when the self-verification test is complete. The Hardware
Test screen displays the functions that pass and fail.
4. (Optional) Tap Export to export the results.
5. Tap Close.
Use Smart Help if any hardware functions fail (see "Request technical support with Smart Help"
page
172).
View the last hardware test result
The self verification test checks the instrument hardware functions. You can view the results of the last
self-verification test that was run.
1. In the home screen, tap
The Self-Verification Test screen is displayed.
2. Tap Hardware test.
3. Tap Last test.
The Self-Verification Test screen displays the functions that pass and fail.
Use Smart Help if any hardware functions fail (see "Request technical support with Smart Help"
page
172).
Perform a peripheral test
The peripheral test takes approximately 5 minutes. Tap
test.
The peripheral test checks the following functions:
• Touchscreen accuracy
• Speaker
• Microphone
• Camera
• Proximity sensor
1. In the home screen, tap
The Self Verification Test screen is displayed.
2. Tap Peripheral test.
3. Tap Start.
The peripheral test begins with the touch accuracy test.
174
QuantStudio
(Settings)4Maintenance and service4Self verification test.
(Settings)4Maintenance and service4Self verification test.
6 Pro Real-Time PCR System and QuantStudio
(Close) on any screen to exit the peripheral
7 Pro Real-Time PCR System User Guide