Chapter 4 Edit a plate before starting a run
4
Options to edit the method of a plate file
1
Figure 8 VeriFlex
VeriFlex
™
1
2. Tap the temperature field in the appropriate VeriFlex
3. Enter a temperature, then tap Enter.
The temperatures in adjacent zones must have a difference of ≤5°C.
4. Tap Done.
A stage with VeriFlex
5. In the Method tab, tap
Note: The VeriFlex
Indicates that this is the Method tab where the VeriFlex
1
Indicates that VeriFlex
2
60
QuantStudio
™
Zones and Auto Delta settings
Zones is selected
™
Zones is denoted with a
4Remove4Done to remove the VeriFlex
™
Zones setting is removed from the Method tab.
™
Zones are assigned. Tap to remove the VeriFlex
™
6 Pro Real-Time PCR System and QuantStudio
™
Zone.
icon.
™
Zones.
1
2
™
Zones settings can be removed
™
Zones.
™
7 Pro Real-Time PCR System User Guide