Perform Calibrations - Thermo Scientific Applied Biosystems QuantStudio 6 Pro User Manual

Real-time pcr system
Table of Contents

Advertisement

Chapter 7 Calibrate and verify instrument performance
7
Perform ROI/uniformity, background, dye, and normalization calibrations for the TaqMan

Perform calibrations

1. In the home screen tap
The Calibration screen is displayed.
2. In the Calibration screen, select a calibration type.
ROI/uniformity
Background
Dye
Normalization
Note: You will be prompted to perform a background calibration after a ROI/Uniformity calibration.
3. (System dye calibration only) Select the dye to calibrate.
• FAM
• VIC
• ROX
4. (Normalization only) Select the reporter dye and passive reference to calibrate.
• FAM/ROX
• VIC/ROX
5. Follow the instructions on the screen to start the calibration.
6. Load the TaqMan
See "Load a TaqMan
CAUTION! The instrument should be used by trained operators who have been warned of the
moving parts hazard.
7. Tap Start.
When the ROI/Uniformity run is complete, you will be prompted to perform a background
calibration.
When the Background calibration is complete, the Calibration Status will display. It will indicate
whether the calibration passed or failed.
100
QuantStudio
(Settings)4Maintenance and service4Calibration.
Calibration type
Array Card into the instrument.
Array Card into the instrument"
6 Pro Real-Time PCR System and QuantStudio
ROI and uniformity
Background
System dye
Instrument normalization
on page
Array Card block
Tap
52.
7 Pro Real-Time PCR System User Guide

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Applied Biosystems QuantStudio 6 Pro and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Applied biosystems quantstudio 7 pro

Table of Contents