Chapter 7 Calibrate and verify instrument performance
7
Perform ROI/uniformity, background, dye, and normalization calibrations for the TaqMan
Perform calibrations
1. In the home screen tap
The Calibration screen is displayed.
2. In the Calibration screen, select a calibration type.
ROI/uniformity
Background
Dye
Normalization
Note: You will be prompted to perform a background calibration after a ROI/Uniformity calibration.
3. (System dye calibration only) Select the dye to calibrate.
• FAM
• VIC
• ROX
4. (Normalization only) Select the reporter dye and passive reference to calibrate.
• FAM/ROX
• VIC/ROX
5. Follow the instructions on the screen to start the calibration.
6. Load the TaqMan
See "Load a TaqMan
CAUTION! The instrument should be used by trained operators who have been warned of the
moving parts hazard.
7. Tap Start.
When the ROI/Uniformity run is complete, you will be prompted to perform a background
calibration.
When the Background calibration is complete, the Calibration Status will display. It will indicate
whether the calibration passed or failed.
100
QuantStudio
(Settings)4Maintenance and service4Calibration.
Calibration type
™
Array Card into the instrument.
™
Array Card into the instrument"
™
6 Pro Real-Time PCR System and QuantStudio
™
ROI and uniformity
Background
System dye
Instrument normalization
on page
™
Array Card block
Tap
52.
7 Pro Real-Time PCR System User Guide