8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
3. Repeat step 2 (beginning with
segments and power levels of your analyzer. Refer to
segment values.
Table 2-13 Frequency Segment Values
Segment
All Models
All Models
All Models
8719ET/ES
8720ET/ES
8722ET/ES
8722ET/ES
4. When all of the frequency segments have been entered, press the following:
DONE
LIST FREQ
Test Setup
5. To set up the PORT 1 measurement (REFLECTION measurement on ET Models), press
the following:
For ES Models:
Trans: FWD S21 (B/R)
Meas
Format
LIN MAG
IF BW
Avg
10
Scale Ref
2
For ET Models:
TRANSMISSN
Meas
LIN MAG
Format
IF BW
Avg
10
Scale Ref
2
Service Guide
) to enter the values for the remaining frequency
ADD
1
2
839.999899 MHz
3
4
4
4
5
ALL SEGS SWEEP
x1
x1
x1
x1
System Verification and Performance Tests
Table 2-13
Start
Stop
Frequency
Frequency
50 MHz
50.000101 MHz
840 MHz
0.84 GHz
8 GHz
8 GHz
13.51 GHz
8 GHz
20.05 GHz
8 GHz
20 GHz
20 GHz
40.05 GHz
Performance Tests
for the frequency
Number
of Points
11
11
201
101
151
151
201
2- 43