Adjustments and Correction Constants

Raw Offset Correction Constants

Raw Offset Correction Constants
This procedure sets the raw offsets for the A,B, and R channels, in preparation for
performing the
"Sampler Calibration Correction Constants (Test 51)," on page
Required Equipment and Tools
Equipment Description
Power Meter
Power Sensor
Adapter 3.5-mm (f) to 3.5-mm (f)
Adapter 2.4-mm (f) to 2.4-mm (f)
RF Cable Set
RF Open
Analyzer and power meter warmup time: 30 minutes.
1. Make sure the A7 Switch is in the alter position (ALT).
2. Zero and calibrate the power meter.
3. Connect the equipment as shown in
3-48
8719ET/ES
8720ET/ES
436A, 437B, 438A, E4418A (EPM-441A) or
E4419A (EPM-442A)
8485A
part number
85052-60012
(part of 85052B)
N/A
85131F
part number
85052-60009
(part of 85052B)
Figure
3-13.
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
3-58.
8722ET/ES
8487A
N/A
part number
85056-60006
(part of 85056A)
85133F
part number
85056-60023
(part of 85056A)
Service Guide