Service Key Menus and Error Messages
GPIB Service Mnemonic Definitions
Analog Bus Codes
ANAI[D]
OUTPCNTR
OUTPERRO
OUTPTESS
TST?
10-32
Measures and displays the analog input. The preset state
input to the analog bus is the rear panel AUX IN. The other
22 nodes may be selected with
enabled (ANABon).
Outputs the counter's frequency data.
Reads any prompt message sent to the error queue by a
service routine.
Outputs the integer status of the test most recently
executed. Status codes are those listed under "TST?".
Executes the power-on self test (internal test 1) and outputs
an integer test status. Status codes are as follows:
0 = pass
1 = fail
2 = in progress
3 = not available
4 = not done
5 = done
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
D
only if the ABUS is
Service Guide