Adjustments and Correction Constants
Protected Option Numbers Correction Constants (Test 50)
6. Press the following:
SERVICE MENU
System
EXECUTE TEST
50
x1
7. Observe the analyzer for the results of the adjustment routine:
• If the analyzer displays Option Cor DONE, you have completed this procedure.
8. If no more correction constant routines are going to be performed, return the A7 Switch
to the NRM position and perform the
In Case of Difficulty
• If the analyzer displays Option Cor FAIL, check the keyword used in step 4 and make
sure it is correct. Pay special attention to the letters "I" or "O", the numbers "1" or "0".
Repeat this entire adjustment test.
• If the analyzer continues to fail the adjustment routine, refer to
Control Troubleshooting,"
office.
3-14
TESTS
YES
"EEPROM Backup Disk Procedure," on page
or contact the nearest Agilent Technologies sales or service
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
3-33.
Chapter 6 , "Digital
Service Guide