PEEK/POKE Menu

To access this menu, press
PEEK/POKE
CAUTION
The PEEK/POKE capability is intended for service use only.
PEEK/POKE ADDRESS
(PEEL[D])
(PEEK)
PEEK
(POKE[D])
POKE
RESET MEMORY
Service Guide
SYSTEM
SERVICE MENU
Allows you to edit the content of one or more memory
addresses. The keys are described below.
accesses any memory address and shows it in the active
entry area of the display. Use the front panel knob, entry
keys, or step keys to enter the memory address of interest.
Displays the data at the accessed memory address.
allows you to change the data at the memory address
accessed by the
front panel knob, entry keys, or step keys to change the
data. The A7CC jumper must be in the "ALT" position in
order to poke.
Resets or clears the memory where instrument states are
stored. To do this, press