Agilent Technologies 8719ET Service Manual page 70

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

System Verification and Performance Tests
Performance Test Record
Power Level and Model Type
P
1
8719ET/ES, 8720ET/ES
8719ES Opt. 007,
8720ES Opt. 007
8722ET/ES, 8722ES Opt. 007
Freq. range (0.05
20 GHz)
8722ET/ES, 8722ES Opt. 007
Freq. range (20
40 GHz)
P
2
8719ET/ES, 8720ET/ES
8719ES Opt. 007,
8720ES Opt. 007
8722ET/ES, 8722ES Opt. 007
Freq. range (0.05
40 GHz)
P
3
8719ET/ES, 8720ET/ES
8719ES Opt. 007,
8720ES Opt. 007
8722ET/ES, 8722ES Opt. 007
Freq. range (0.05
20 GHz)
P
4
8719ET/ES, 8720ET/ES
8719ES Opt. 007,
8720ES Opt. 007
2-48
Source Linearity
Results
Minimum
Specification
Min.
−0.35 dB
_______
−0.35 dB
_______
−0.60 dB
_______
−0.60 dB
_______
−0.60 dB
_______
−0.35 dB
_______
−0.35 dB
_______
−1.0 dB
_______
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Maximum
Specification
Max.
_______
0.35 dB
_______
0.35 dB
_______
0.60 dB
_______
0.60 dB
_______
0.60 dB
_______
0.35 dB
_______
0.35 dB
_______
1.0 dB
Measurement
Uncertainty
±
0.03 dB
±
0.03 dB
±
0.03 dB
±
0.04 dB
±
0.04 dB
±
0.03 dB
±
0.03 dB
±
0.04 dB