Front Panel Troubleshooting (A1/A2); Check Front Panel Leds After Preset - Agilent Technologies 8719ET Service Manual

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers

Front Panel Troubleshooting (A1/A2)

Check Front Panel LEDs After Preset

1. Press
on the analyzer.
Preset
2. Observe that all front panel LEDs turn on and, within five seconds after releasing
, all but the CHAN and TRANS FWD and REFL FWD LEDs turn off.
Preset
• If all the front panel LEDs either stay on or off, there is a control problem between
the CPU board (A7) and the front panel set (A1/A2). See
page
6-16.
• If you suspect that one or more LEDs are burned out, then replace the keypad
assembly (A1).
NOTE
TRANS FWD and REFL FWD LEDs, and TRANS REV and REFL REV LEDs
problems may be caused by the malfunction of the LED board or the transfer
switch
• If, at the end of the turn on sequence, the channel 1 LED is not on and all GPIB
status LEDs are not off, continue with
Service Guide
Front Panel Troubleshooting (A1/A2)
"Inspect Cables" on
"Identify the Stuck Key" on page
Digital Control Troubleshooting
6-14.
6- 13