Adjustments and Correction Constants
Output Power Adjustments

Adjusting the Mid Band Power

1. Connect the equipment as shown in
Figure 3-4 Setup for Output Power Adjustments
Table 3-3 Output Power Adjustment Values
Model/Option
8719ES and 8720ES
8719ET and 8720ET
8719ES Option 007
8720ES Option 007
8722ES
8722ET
8722ES Option 007
2. Press the following:
Preset
POWER
Sweep Setup
PWR RANGE MAN
Enter the value from the "Power A" column in
3-20
Figure
Power A
(dBm)
−3
+2
−13
−8
POWER RANGES
RANGE 0
3-4.
Power B
(dBm)
0
+5
−10
−5
Table
3-3. Press
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
.
x1
Service Guide