System Verification and Performance Tests
Performance Tests
Frequency Accuracy Performance Test
This test checks the frequency accuracy of the analyzer at or near its maximum frequency.
Table 2-6 Required Equipment
Equipment Description
Frequency Counter
RF Cable Set
Analyzer warmup time: 30 minutes.
NOTE
On the frequency counter, set the resolution to 1kHz or less.
1. Connect the equipment as shown in
PRESET: FACTORY
Preset
Sweep Setup
CW FREQ
Figure 2-17 Frequency Range and Accuracy Test Setup
2. Select the analyzer CW frequency:
• Agilent 8719ET/ES, press
• Agilent 8720ET/ES, press
• Agilent 8722ET/ES, press
3. Record the frequency counter reading for each frequency point in the results section on
the "Frequency Accuracy" performance test record at the end of this chapter.
2-34
8719ET/ES
8720ET/ES
53150A or 5350B
53151A
85131F
85133F
Figure 2-17
and press the following:
Preset
13.5
G/n
20
G/n
26
G/n
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
8722ET/ES
or
5351B
Service Guide