8719ET/20ET/22ET
Theory of Operation
8719ES/20ES/22ES
System Operation
RF Network Analyzers
System Operation
The 8719ET/ES, 8720ET/ES, and 8722ET/ES microwave network analyzers integrate a
synthesized source, signal separation devices, a three or four channel receiver for
measurement of test device characteristics, and a large-screen display.
Figure 12-1
is a
simplified block diagram of the network analyzer system.
Figure 12-1 Simplified System Block Diagram
The built-in synthesized source of the analyzer generates a swept or continuous wave (CW)
signal in the following ranges:
8719ET/ES
8720ET/ES
8722ET/ES
50 MHz to 13.51 GHz
50 MHz to 20.05 GHz
50 MHz to 40.05 GHz
The source output power is leveled by an internal ALC (automatic leveling control) circuit.
The output maximum power levels at the front panel are listed in
Table 12-1 on page
12-4.
A portion of the source signal is routed to the R sampler receiver, and fed back to the
source for phase lock.
Service Guide
12- 3