Performing The Verification Tests - Agilent Technologies 8719ET Service Manual

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

System Verification and Performance Tests
System Verification

Performing the Verification Tests

1. When prompted, insert the 20 dB attenuator (Ver Dev 1) as shown in
Figure 2-11 Verification Device Connections
2. To start the tests, press
3. If the record function is off (printout is not required), the program will pause after each
S-parameter measurement and you will need to press
measurement.
NOTE
For ES models, there are eight measurements (magnitude and phase for each
of the four S-parameters for each verification device).
For ET models, there are four measurements (magnitude and phase for each
forward S-parameter for each verification device).
2-26
.
CONTINUE
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Figure
after each
CONTINUE
Service Guide
2-11.

Advertisement

Table of Contents
loading

This manual is also suitable for:

8720et8722et8719es8720es8722es

Table of Contents