Receiver Troubleshooting
Information on This Chapter
Information on This Chapter
Use this procedure only if you have read
the procedures in the order given, unless instructed otherwise.
This section can be used to determine which receiver assembly of the instrument is faulty.
The two receiver assemblies that affect all three signal paths are the digital IF (A10) and
reference assemblies (A12). The receiver assemblies that are associated with specific
signal paths are listed in
Table 8-1 Receiver Assemblies and Associated Paths
Signal Path
Port
R or R1
R2 (Option 400)
A
B

Assembly Replacement Sequence

The following steps show the sequence to replace an assembly in the network analyzer.
Step 1. Identify the faulty group. Begin with
Follow up with the appropriate troubleshooting chapter that identifies the faulty
assembly.
Step 2. Order a replacement assembly. Refer to
Step 3. Replace the faulty assembly and determine what adjustments are necessary.
Refer to
Chapter 14 , "Assembly Replacement and Post-Repair Procedures."
Step 4. Perform the necessary adjustments. Refer to
Correction Constants."
Step 5. Perform the necessary performance tests. Refer to
Verification and Performance Tests."
8-2
Chapter 4 , "Start Troubleshooting Here."
Table
8-1.
Directional
Coupler
internal
NA
internal
NA
1
A62
2
A63
Sampler
2nd
Assembly
Converter
A64
A6
A67
A6
A65
A4
A66
A5
Chapter 4 , "Start Troubleshooting Here."