Start Troubleshooting Here
Step 3. GPIB Systems Check
Step 3. GPIB Systems Check
Check the analyzer's GPIB functions with a known working passive peripheral (such as a
plotter, printer, or disk drive).
1. Connect the peripheral to the analyzer using a known good GPIB cable.
2. Press
SYSTEM CONTROLLER
Local
peripheral.
3. Then press
SET ADDRESSES
addresses will be recognized by the analyzer. The factory default addresses are:
Table 4-2 Factory Set GPIB Addresses
Device
All Analyzers
Plotter port
Printer port
Disk (external)
Controller
Power meter
NOTE
You may use other addresses with two provisions:
• Each device must have its own address.
• The address set on each device must match the one recognized by the
analyzer (and displayed).
Peripheral addresses are often set with a rear panel switch. Refer to the
manual of the peripheral to read or change its address.
4-10
to enable the analyzer to control the
and the appropriate softkeys to verify that the device
GPIB Address
16
5
1
0
21
13
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Service Guide