Checking The R Channel Output (Et And Es Models) - Agilent Technologies 8719ET Service Manual

Network analyzer
Hide thumbs Also See for 8719ET:
Table of Contents

Advertisement

Start Troubleshooting Here
Receiver Check

Checking the R Channel Output (ET and ES Models)

1. Connect the equipment as shown in
Figure 4-7 Setup for the R Channel Receiver Check (ES and ET Models)
2. Zero and calibrate the power meter. Set the calibration factor to the 1 GHz value. (See
the power meter manual for instructions on setting the calibration factor.)
3. Press the following:
Preset
PRESET: FACTORY
4. Press the following:
Meas
INPUTPORTS
CW FREQ
Sweep Setup
5. Calculate the difference between the power meter reading and the marker reading on
the analyzer display. This power difference is the frequency response of the R channel.
6. Repeat steps 4 and 5 at the frequency points where a problem is suspected or check
different frequency points across the band (set the calibration factor on the power meter
for each frequency measured). For the characteristic values of the frequency response,
see
Table 4-3 on page
4-22
Figure
Preset
R
Marker
4-21.
4-5.
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Service Guide