Start Troubleshooting Here
Step 2. Operator's Check

Procedure to Perform Operator's Check

Analyzer warmup time: 30 minutes
PORT 1 or REFLECTION port check
1. Connect a short to PORT 1 (REFLECTION port on ET models).
2. Press the following:
Preset
PRESET: FACTORY
SERVICE MENU
System
EXECUTE TEST CONTINUE
• If the message PRESS[CONTINUE] appears on the display during the test, that
particular attenuator setting check has failed. Press
other attenuator settings.
• If the message FAIL appears on the analyzer display, the analyzer has failed the
PORT 1 operation check. Refer to the flow charts on
• If the message DONE appears on the analyzer display, the analyzer has passed the
PORT 1 operation check.
PORT 2 Check (ES Models)
3. Connect the short to PORT 2.
4. Press the following:
Preset
PRESET: FACTORY
SERVICE MENU
System
22
x1
EXECUTE TEST CONTINUE
• If the message PRESS[CONTINUE] appears on the display during the test, that
particular attenuator setting check has failed. Press
other attenuator settings.
• If the message FAIL appears on the analyzer display, the analyzer has failed the
PORT 2 operation check. Refer to the flow charts on
• If the message DONE appears on the analyzer display, the analyzer has passed the
PORT 2 operation check.
4-8
Preset
TESTS EXTERNAL TESTS
Preset
TESTS
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
to check the
CONTINUE
page 2-4
and
page
2-5.
to check the
CONTINUE
page 2-4
and
page
2-5.
Service Guide